Reprocessed Agilis Steerable Introducer Sheath by Sterilmed- Class I Recall – Improper Seal of Sheath Hub:

Reprocessed Agilis Steerable Introducer Sheath by Sterilmed- Class I Recall – Improper Seal of Sheath Hub:

[Posted 01/02/2018]

AUDIENCE: Risk Manager, Cardiology

ISSUE: The Agilis Steerable Introducer Sheath’s hemostatic valve, which prevents blood from flowing back through the valve, may fail due to an improper seal of the sheath hub. Improper seals can allow blood to leak through the hub, cause the cap to fall off during the procedure, or can create a difference in pressure that allows air into the circulatory system (air embolism).

  • Model numbers and lot numbers: STJ408309, STJ408310, STJG408324; All product lots
  • Manufacturing and Distribution Dates: January 1, 2017 to May 5, 2017

The use of affected products may cause serious health consequences for patients, including death.

BACKGROUND: The Agilis Steerable Introducer Sheath is used to insert and position various cardiovascular catheters in the heart, including on the left side of the heart through the wall of tissue that separates the right and left chambers of the heart (interatrial septum).

The improper seal occurs when not enough glue is used to reattach the cap to the hub after reprocessing. Too much glue can also block the sheath valve and make the device unusable.

RECOMMENDATION: Sterilmed has provided the following instructions for health care facilities and providers:

  • Examine inventory immediately to determine if you have this product.
  • Do not use any of the affected products, and return any unused product to the company.
  • Share this information with the appropriate staff at your facility.
  • Continue to monitor patients treated with the Sterilmed Reprocessed Agilis Steerable Introducer Sheath as normal.

Healthcare professionals and patients are encouraged to report adverse events or side effects related to the use of these products to the FDA’s MedWatch Safety Information and Adverse Event Reporting Program:

  • Complete and submit the report Online: www.fda.gov/MedWatch/report
  • Download form or call 1-800-332-1088 to request a reporting form, then complete and return to the address on the pre-addressed form, or submit by fax to 1-800-FDA-0178

Recalled Product

  • Name: Sterilmed Reprocessed Agilis Steerable Introducer Sheath
  • Product codes: PNE
  • Model numbers and lot numbers: STJ408309, STJ408310, STJG408324; All product lots
  • Manufacturing and Distribution Dates: January 1, 2017 to May 5, 2017
  • Number of Affected Devices: 112

Device Use

The Agilis Steerable Introducer Sheath is used to insert and position various cardiovascular catheters in the heart, including on the left side of the heart through the wall of tissue that separates the right and left chambers of the heart (interatrial septum).

Image of the device.

Reason for Recall

The Agilis Steerable Introducer Sheath’s hemostatic valve, which prevents blood from flowing back through the valve, may fail due to an improper seal of the sheath hub. Improper seals can allow blood to leak through the hub, cause the cap to fall off during the procedure, or can create a difference in pressure that allows air into the circulatory system (air embolism).

The improper seal occurs when not enough glue is used to reattach the cap to the hub after reprocessing. Too much glue can also block the sheath valve and make the device unusable.

The use of affected products may cause serious health consequences for patients, including death.

Who May be Affected

Hospital clinicians using these devices for cardiac catheterization procedures.

Patients undergoing cardiac catheterization procedures. Patients with a lower body mass index (BMI) may be more at risk if blood loss occurs. Smaller patients and patients with pre-existing decreased pulmonary reserve may be more susceptible to air embolism.

What to Do

Sterilmed has provided the following instructions for health care facilities and providers:

  • Examine inventory immediately to determine if you have this product.
  • Do not use any of the affected products, and return any unused product to the company.
  • Share this information with the appropriate staff at your facility.
  • Continue to monitor patients treated with the Sterilmed Reprocessed Agilis Steerable Introducer Sheath as normal.

Contact Information

Sterilmed, Inc.
5010 Cheshire Lane
Plymouth, MN 55446
Phone: 888-541-0078
Email: sterilmedcs@its.jnj.com

Date Recall Initiated

June 12, 2017

Advertisements

Leave a Reply

Fill in your details below or click an icon to log in:

WordPress.com Logo

You are commenting using your WordPress.com account. Log Out /  Change )

Google+ photo

You are commenting using your Google+ account. Log Out /  Change )

Twitter picture

You are commenting using your Twitter account. Log Out /  Change )

Facebook photo

You are commenting using your Facebook account. Log Out /  Change )

Connecting to %s